Set
the timer to the desired cleaning time.
Heavily soiled instruments should be cleaned for at
least one ten minute cycle.
Lightly soiled cannulated instruments or standard hand
instruments can usually be cleaned in about 5 minutes.
The cleaning process will begin.
CAREFULLY
open the lid to make sure you have proper flow through
the instruments and all connections are secure.
Never open the unit unless you are wearing approved
eye protection and clothing.
Increase or decrease the flow as needed using the Flow
Control adjustment on the front panel of the unit. Return
the lid to the closed position. Never allow the unit
to run with the lid in the open position.
REMEMBER:
A higher volume of irrigation does not necessarily mean
cleaner instruments. The faster the water moves through
the instruments, the less energy from the ultrasonics
can be achieved. Adjust the flow to maintain a slow,
steady flow of liquid from the instrument. This will
allow the ultrasonics to perform the physical chore
of debris removal. After you have adjusted the flow
and closed the lid, continue your routine.
To gain the full benefit of the unit, it is suggested
that you have an additional set-up basket to allow the
next load of instruments to be prepared while these
instruments are being cleaned.
The cleaning process can be repeated throughout an 8
hour shift without changing the water or filter in the
basin, unless the water becomes extremely dirty. |